• Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
  • Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
  • Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
  • Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
  • Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
  • Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome

Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome

Customized: Non-Customized
Certification: ISO13485
Scale: Medium
Output: Fully Suspended
Power Supply: 220V or 110V
Working Modes: 5
Customization:
Gold Member Since 2020

Suppliers with verified business licenses

Rating: 5.0/5
Manufacturer/Factory, Trading Company
  • Overview
  • Product Description
  • Detailed Photos
Overview

Basic Info.

Model NO.
CM-350D
Pure Cut
1W~350W (Max. 400W, Load 500 Ω)
Blend Cut 1
1W~250W (Load 500 Ω)
Endu Cut
99% (Load 500 Ω)
Spray Coagulation
1W~80W (Load 500 Ω)
Desicate Coagulation
1W~120W (Load 500 Ω)
Standard Bipolar Coagulation
1W~70W (Load 100 Ω)
Transport Package
Carton
Specification
7.5kg
Trademark
Kindsway
Origin
China
Production Capacity
500units/Month

Product Description

Product Description

CM-350D Electrosurgical Unit 

5 modes
-Pure cut (400 W),
-mixed cut,
-spray set,
-soft set,
-double set (70 W)

Type:
The output of this machine is fully suspended, with two independent and isolated CF type
anti-defibrillation application parts (Monopolar and Bipolar), Non-AP,Non-APG type
common equipment, and 220V ± 10% or 110V ± 10% network power supply.
Function:
The 416KHz (± 10%) high frequency current generated by this machine can perform surgical
operations such as cutting and coagulation of biological tissues, and has monopolar cut:
pure cut, blend cut 1, blend cut 2; Monopolar coagulation: spray coagulation, soft
coagulation and standard bipolar six working modes.
Full-function automatic power compensation.
The high-frequency generator has a universal endoscope interface, which is used with
gastroscope, enteroscope, bronchoscope, laparoscope and other endoscopes.
Support bipolar closure and separation to meet the requirements of surgical bipolar
treatment without sticking the negative plate.
Patient circuit bipolar electrode plate contact quality detection system.
Monopolar output can be controlled either by hand or foot, and monopolar/bipolar
automatic output.
* This machine has a power-off protection circuit, which can memorize the output set
value of each function in real time.
Remote fault diagnosis.
* Human voice alarm function
Passed EMC test.
Rated output power:
Pure cut:1W~350W (Max. 400W, load 500 Ω);
A high-frequency alternating current yields smooth, rapid cuts that evoke little to know
hemostasis. (No-focus precision cutting, 100% cutting.)
Blend cut 1:1W~250W (load 500 Ω);
Preset blended cut modes give the surgeon varying degrees of hemostasis in cut mode.
(Superficial focal gamma accurate cutting 80% cutting, 20% coagulation.)
ENDU Cut: 99% (load 500 Ω);
The mode is used to automatically controlled cut system to reduce the complication
rate of endoscopic sphincterotomy (EST) and serum hyperamylasemia after EST
compared to the conventional blended cut mode.
Spray coagulation:1W~80W (load 500 Ω);
Spray output is made for direct coagulation at the tissue with spark. Above spark does
not give any cutting effect as its CREST FACTOR is high. (Massive coagulation)
Desicate coagulation:1W~120W (load 500 Ω);
Coagulation in which the active electrode is in direct contact with the tissue is referred
to as desiccation and is the type of coagulation used in most surgeries it ensures
pinpoint desiccation with less destruction of peripheral tissue.
Standard bipolar coagulation:1W~70W (load 100 Ω);
BI-COAG - For highly precise control over coagulation process and privations of
carbonization.
Power consumption of the whole machine: ≤ 1100VA. (cut function 350W);
Normal operation
Ambient temperature range: 5 ºC~40 ºC;
Relative humidity range: ≤ 80%;
Atmospheric pressure range: 86.0~106.0kPa;
Power supply: 220V±10%, 50±1Hz; or 110V±10%, 60±1Hz
Working frequency: 416kHz;
Transportation and storage
Ambient temperature range: - 40 ºC~55 ºC;
Relative humidity range: ≤ 95%;
Atmospheric pressure range: 50.0~106.0kPa.

Configuration List:
 
Main unit 1 PC
Disposable hand switch 2PCS
Disposable neutral plate 3PCS
Disposable neutral plate cable 1 PC
Bipolar forceps 1 PC
Bipolar forceps cable 1 PC
Double foot switch 1 PC
Power cable 1 PC
GBXP5*20(5A) Fuse 2PCS
USER'S MANUAL       1 PC
WARRANTY CARD 1 PC
 
Detailed Photos

Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
Cm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife ElectrotomeCm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife ElectrotomeCm-350d Electrosurgical Unit Surgical Instrument Equipment Electricity Knife Electrotome
 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Gold Member Since 2020

Suppliers with verified business licenses

Rating: 5.0/5
Manufacturer/Factory, Trading Company
Registered Capital
1000000 RMB
Plant Area
<100 square meters