MS400 Multi Parameter Simulator ECG/EKG Patient Simulator-Touch Screen-CE Certificate
Portable ECG/EKG simulatoris a portable device, it can simulate accurately 12-lead ECG, Respiration, Temperature, 2-channel IBP or 4-channel IBP.
This Portable ECG/EKG simulator can simulate the following item
1.35 kinds of arrhythmia
2.4-channel IBP data.
3.Respiration signal with adjustable respiration rate, impedance and baseline impedance, and 3 kinds of apnea data.
4.4 kinds of temperature data.
5.Square wave, triangle wave, pulse wave and sine wave with adjustable frequency.
6.Standard ECG calibration, adult and pediatric ECG signal.
-ECG waveform complied with the standard of ANSI/AAMI EC13-2002, AHA, MIT ECG database, other ECG database and user-defined waveform data stored in TF card can be displayed by the simulator and exported to the equipments to be tested.
-Each waveform can be displayed by sliding the 3.5'' TFT touch screen, easy to operate.
-Power supply: Rechargeable lithium battery or linear adapter, which reduces interference.
Parameters: ECG
1.Conventional: All leads have the same voltage with the II-lead as the reference
2.Amplitude: 0.5, 1.0, 1.5 and 2.0mV
3.ECG Rate: 15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300, 320, 340 and 350 BPM
4.ECG artifacts: 50 and 60 Hz, muscle, baseline and breathing
5.The elevation of the ST segment: +0.8 to -0.8 mV, the step is 0.1 mV
Arrhythmia
1.Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial Tachycardia, 2.Nodal Rhythm, Supraventricular Tachycardia
3.Ventricular: Vent Fib (coarse), Vent Fib (fine), PVCs 6/min, PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5 PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
Premature Beat: Atrial PAC, Nodal PNC, PVC1 LV Focus, PVC1 E LV Focus, R on T LV Focus, PVC2 RV Focus, PVC2 E RV Focus, R on T RV Focus, Multifocal PVC
4.Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block
4 IBP Channels
1.The default: 00/00mmHg
2.Dynamic IBP(unit: MmHg)
-Channel 1:
Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Ventricle: 25/00
-Channel 2:
Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
-Channel 3:
Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
-Channel 4:
Auto Swan-Ganz (every 15 seconds): RA(CVP), RV, PA, PAW (Pulmonary Wedge)
Man Swan-Ganz: Press ENTER key to change every time
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg
Channel 3: 0, 20, 40, 60, 80 and 100mmHg
Channel 4: P4 tracks the static pressure levels of P3
Temperature
Four temperature: 0º C(32° F), 24º C( 75° F), 37º C(98.6° F), 40º C(104° F)
Accessories
Standard accessories 9V/1A(CE) power adapter, an User manual, a packaging box, a plastic packaging.
Optional accessories a power adapter(15V/1A)(UL), a power adapter(15V/1A), a temperature cable, an IBP cable, a CD disk(including a case conversion software)