Medical air compressor Kindsway-3000
Integrated with American GAST/THOMAS oil-free pump, multiple well-known pneumatic components such as Japanese SMC& UK NORGREN, resulting in further savings on operating and maintenance costs
Description
Operation
Production Show
Clinic Application
Compressed air pump produces compressed air with high temperature and high pressure. The
compressed air will be cooled down by cooling system. The water trap will drain off the moist in
compressed air, and the pressure regulator will reduce the pressure of compressed air to operating
pressure then finally the air can be supplied through the outlet.
3. Technical Data
Outlet flow at pressure 3-4bar 60L/min
Peak flow 120L/min
Voltage/Frequency 220V/ 50Hz 110V/50-60Hz
Outlet connection NIST
Sound level dB(A) 53-55dB(A)
Mode of operation Continuously
Separation of condensed water Automatic
Outlet pressure 0.3 Mpa - 0.4 Mpa (3 bar - 4 bar)
Air tank capacity 0.5L
Operating pressure of safety valve 7 bar
Adjustment of outlet air pressure Pressure regulator
Dimensions of compressor w*d*h 580 × 580 × 480 mm
Dimensions of packaging w*d*h 700 × 700 × 580 mm
Net weight 52KG
Gross weight 55KG
A. User should remove the package and check the exterior of the compressor to find if there is any
damage in the exterior.
B. Check the relevant parts to find if they are complete .
C. Connect power supply and test functions of the compressor.
D. The ground where the Medical Air Compressor is going to be operated should be flat.
E. The stabilizing screws for transport should be removed and replaced by two plugs configured before
using.
F. Power supply cord should be well grounded to avoid damage to user under abnormal conditions.
G. The following requirements of power socket should be satisfied to avoid connect problems and fire
hazards.
a. Operating electric current state on the power supply should correspond to the main power
input as described in this document.
b. With perfect function.
H. The rated value of power cable should be adapted to the maximum operating current of the
compressor to avoid accident caused by overheat power wires after long time use.